| Introducing the |
 |
LiDCO™plus is a Hemodynamic Monitor, providing continuous,
reliable and accurate assessment of the hemodynamic status of critical
care and surgery patients. This is achieved by running two proprietary
algorithms: a continuous arterial waveform analysis system (PulseCO™)
coupled to a single point lithium indicator dilution calibration system
(LiDCO™).
The developmental objective of the LiDCO™plus Hemodynamic
Monitor was to develop a novel platform monitor that would provide an
interpretable user interface displaying real time: blood pressure, pre
load, cardiac output/oxygen delivery and after load parameters.
Overview of the LiDCO™plus Hemodynamic Monitor Technology
The advent of a reliable method for beat-by-beat derivation of hemodynamic
parameters from the existing arterial pressure catheter and waveform provides
surgeons, anesthesiologists and other intensivists a significantly improved
hemodynamic monitoring option. The LiDCO™plus Hemodynamic
Monitor is designed to improve patient management. The LiDCO™plus
Hemodynamic Monitor requires calibration with the absolute value obtained
from a single point cardiac output measurement once every eight hours.
The PulseCO™ Software
The PulseCO™ software provides continuous assessment of a patient's
hemodynamic status, by analysing and processing the arterial pressure
signal obtained from the primary blood pressure monitor.
The concept of using the blood pressure waveform to measure blood flow
changes has been previously suggested by a number of investigators. Such
methods provide cardiac output trend data following calibration by an
independent, preferably an indicator dilution based, cardiac output measurement.
|
|
|
PulseCO™ System Autocorrelation Algorithm
Step 1 - Arterial Pressure Transformation into a Volume Time Waveform
An accurate way of determining the change in blood volume in
the arterial tree from maximum dilatation to minimum dilatation
would allow an estimate of the volume of blood flowing out of
the arterial tree during a period slightly longer than diastole.
Since the whole period of the cardiac cycle usually bears a
(near) fixed relation to diastole simple scaling up would give
the stroke volume. The relationship between the capacity of
the arterial side of the circulation and the intravascular pressure
can be expressed as the compliance: pressure change per unit
volume change. All would be relatively simple if this was constant.
However, arterial compliance has been shown to change as arterial
pressure changes. A stiffening of the vasculature occurs as
pressure and volume increase such that, at higher pressures,
a given increase in pressure expands the arterial tree by a
smaller volume. Nevertheless, the form of this curvilinear relationship
(approximately exponential), though differing in its scaling,
appears to be very similar in different subjects. Using a lookup
table the pressure waveform can be used as the basis for calculating
a continuous curve describing the general form of the arterial
volume changes with every cardiac cycle for which arterial blood
pressure is available.

Step 2 - Deriving Nominal Stroke Volume and the Heartbeat Duration
In order to obtain cardiac output (volume per unit time) we
require stroke volume (or at least a value proportional to it
ie nominal stroke volume) and also the duration of the cardiac
cycle to calculate flow. The autocorrelation technique gives
us both.
Nominal Stroke Volume: First of all the software subtracts
the mean value of the derived arterial blood volume record,
giving a description of how much the arterial blood volume changes
around it. This is periodic like a sine wave but with different
shaped areas above and below zero. The figure below shows how
the method works by first using a pure sine wave and subjecting
it to the procedure. We firstly obtain an estimate of the mean
deviation from zero by multiplying all values of the waveform
by themselves. This gives positive waves for both the positive
and negative parts of the original sine wave - effectively a
double waveform. The mean of the values of the new (double)
waveform is otherwise known as the mean square. The square root
of this value is a constant proportion of the amplitude of the
original waveform known as the root mean square, or RMS value.
It is approximately 0.7 of the amplitude. The original sine
wave and the squared (double) waveform are shown for three cycles.

Estimation of the Heart Beat Duration: The precise period
of the cycle can be obtained by moving one version of the volume
waveform successively, step by step, relative to another. Cross
multiplication and addition of the answers now gives values,
which are both positive and negative. The sum for a given displacement
(often referred to as a tau shift) becomes less, and with maximum
opposition of the two versions of the record a negative mean
square difference is obtained. This is not quite as large as
the positive version originally obtained because of the asymmetrical
nature of the waveform. The sum of the two magnitudes (maximum
reinforcement at tau 0 and maximum opposition at some intermediate
point in the cycle - not precisely half way) gives a shifted
times unshifted value very close to the same squared value for
zero tau shift.
Continuation of the step by step movement of one version of
the waveform relative to the other eventually causes positive
reinforcement again, when the moving waveform arrives one cycle
further along relative to the other. This process does not give
a very close value to the mean square volume: the two versions
tend to vary sufficiently once they are one cycle shifted relative
to each other. However, the second positive peak in the autocorrelogram
occurs at a tau shift, which represents the duration of the
cardiac cycle.
Step 3 - Nominal Stroke Volume & Calibration
With an estimate of the square of the volume draining from
the arteries as they shrink from their maximum to their minimum
size we have a volume linearly related to the stroke volume
ie the nominal stroke volume. With the period of the heartbeat
defined precisely by the interval to the first peak in the autocorrelogram
we can therefore calculate a value linearly related to the cardiac
output following calibration with an indicator dilution system.
The nominal stroke volume and cardiac output are initially uncalibrated.
They are converted to calibrated data ie 'true stroke volume/cardiac
output' by multiplying the nominal stroke volume by a calibration
factor (patient specific correction factor [CF].) The patient
specific CF is determined by inputting the cardiac output derived
from an independent method of deriving cardiac output ie LiDCO™
System.
Ventricular Preload and Fluid Responsiveness
Introduction
Assessing the vascular fluid volume status of ventilated patients
and their likely response to additional fluid administration is
vitally important in the management of critically ill patients.
Most patients will require some degree of fluid resuscitation and
subsequent fluid maintenance. There is a clear requirement for a
non or minimally invasive method of accurately assessing fluid status.
Adequate fluid volume is important because of the observation that
the degree to which the ventricle is filled before contraction is
itself the main determinant of the volume of blood ejected. In other
words the heart performs more efficiently when appropriately filled.
The term preload refers to maximum stretch on the heart's muscle
fibres at the end of diastolic filling. The degree of stretch is
determined by the volume of blood contained in the ventricle at
that time. The greater the volume of blood in the ventricles, the
greater the stretch of the myocardial muscle fibres. Increased stretch
results in a greater contraction and therefore an increased stroke
volume is ejected to the body tissues. This relationship between
ventricular end-diastolic filling volume and stroke volume is known
as Starling's Law of the Heart. This states that the energy of contraction
of the muscle is proportional to the pre contraction length of the
muscle fibre.
A very common requirement in surgery and critical care patients
is the need to rapidly increase the blood flow and hence oxygen
delivery to the body. Given Starling's Law of the Heart the most
commonly used intervention is to increase blood flow through the
infusion of fluids. Unfortunately, this intervention itself is not
without risks that include: volume overload, lung odema and acidosis
from excess chloride ion administration. There are considerable
differences in each patient's response to additional volume. Therefore
there is a requirement for a hemodynamic parameter(s) that would
predict the likely response to increased fluid volume and then to
carefully monitor the actual stroke volume response as fluids are
infused.
Unfortunately a convenient real time means of measuring the myocardial
muscle fibre length/stretch response of the ventricle is not, at
present, available for most critical care patients. The most common
method used involves the use of a pulmonary artery catheter based
measurement system. Parameters measured have included pulmonary
artery occlusion pressure, and right or left ventricular end-diastolic
volume. These measurements are by necessity restricted to a small
number of critical care patients as the procedure requires the insertion
of a catheter into the pulmonary artery. Furthermore these measurements
have been reported as having variable success in fulfilling the
requirements of estimating preload, ventricular volumes and subsequent
response of the stroke volume to a fluid challenge.
Arterial pressure waveform derived parameters as indicators of preload status and response to volume infusion
Mechanical ventilation induces cyclical changes in ventricular
stroke volume. Stroke volume is altered by the transient increase
of intrathoracic pressure, after load and lung volume that occurs
on inspiration. These collectively decrease the right ventricular
filling volume and the stroke volume that is ejected into the pulmonary
artery. Following a phase lag of two to three heart beats, representing
the time taken for the blood to transit from the pulmonary circulation
back to the left heart, this reduced ejection volume in turn leads
to diminished filling of, and stroke volume from, the left ventricle.
This variation in stroke volume also results in a cyclical fluctuation
of arterial blood pressure (Systolic Pressure Variation SPV and
Pulse Pressure Variation PPV.) The importance of these continuous
measurements are that the magnitude of the respiratory changes in
left ventricular stroke volume, or arterial pressure changes, should
be proportional to the degree of ventricular preload dependence.
In other words the patient will show greater cyclical changes in
these parameters when the ventricle is operating on the steep, rather
than flat portion of the Starling curve. These parameters can therefore
provide an indication of the patient's preload status and likely
response to the administration of further volume. By superimposing
the "Blood Pressure Waveform" window onto the Trend, Graph or Chart
screens of the LiDCO™plus, a continuous measure of stroke
volume variation (SVV) and the other volume status indicators of
systolic pressure and pulse pressure variation are displayed.
Systolic Pressure Variation (SPV)
is the difference between the maximum and minimum values of systolic arterial pressure recorded
over a single respiratory cycle.
Pulse Pressure Variation (PPV%) is measured over a single
respiratory cycle and defined as the maximal pulse pressure (systolic
- diastolic pressure) less the minimal pulse pressure divided by
the average of these two pressures.
Stroke Volume Variation (SVV%) is measured over a single
respiratory cycle and defined as the maximal stroke volume less
the minimal stroke volume divided by the average of these two stroke
volumes.
Contraindications for use of the PulseCO™ Autocorrelation Algorithm
The performance of the software may be compromised in the following
patient groups:
- Patients with aortic valve regurgitation
- Following aortic reconstruction - a recalibration is required
- Patients being treated with an intra aortic balloon pump
- Patients with highly damped peripheral arterial lines
- Patients with pronounced peripheral arterial vasoconstriction
Analysis of the respiratory changes in the preload parameters of:
SPV, PPV% and SVV% is compromised in patients with cardiac arrythmias
|